A 79-year-old patient with an aortic dissection that was treated by surgical replacement of the ascending aorta in 2006 and a Bentall procedure in 2015, presented for a follow up CT scan of the aorta.
Due to the patient’s impaired renal function, a Deep Learning Spectral CTA scan of whole aorta was performed on an Aquilion ONE / PRISM Edition with 80 ml of 270 mgI/ml contrast media injected at 3 ml/s. This represented a 46% reduction of iodine load compared to the standard site protocol using 100 ml of 400 mgI/ml strength contrast.*
* Optimization of contrast usage is only recommended within the dosing ranges that appear in approved iodinated contrast drug labeling.
Virtual monochromatic images generated at 75 keV demonstrate suboptimal vascular opacification for evaluation of the aorta due to both reduction in iodine concentration and amount of iodine used for this patient. However, the spectral images at 40 keV show excellent improvement in vascular contrast enhancement, particularly within the patent false lumen of this Stanford type B dissection. 3D Global Illumination rendering (GIR) images at 40 keV clearly demonstrate a stable type B dissection (arrow).
Deep Learning Spectral provides excellent low noise virtual monochromatic images (VMI) across the entire keV spectrum. This allows clinicians to fully benefit from the increased contrast enhancement in the low keV range (35- 60 keV) improving vascular assessment without the usual high noise associated with the lower keV levels.
As shown above the density of iodinated contrast increases as the VMI energy level approaches the k-edge of iodine at 33.2 keV.